Cemented total knee arthroplasty was performed on 108 patients 110 knees at seven hospitals in three countries.
Ceramic multigen plus knee with biolox delta ceramic femoral component.
Major prerequisite for the development of the biolox s delta ceramic knee components was the use of existing designs in order to provide the surgeon with a well known surgical technique and the same instruments as for the standard metal femoral component short learning curve.
Its design is identical to that of the multigen plus knee except for having a symmetric anterior surface.
Biolox delta ceramic femoral component integrated in the multigen plus total knee system lima lto san daniele italy is introduced in the european market.
A national prospective duocenter study on the clinical and radiological outcome of the multigen plus total knee system with a biolox delta ceramic femoral component article january 2010 with 17.
The first implantation of the multigen plus knees with the biolox delta ceramic femoral component was successfully performed in our department in december 2006.
The multigen plus total knee system lima corporate villanova di san daniele del friuli italy is a non constrained surface replacement consisting of a symmetric cruciate retaining cr cemented biolox delta ceramic femoral component the cemented metal tial6v4 tibial tray is combined with fixed bearing ultra high molecular weight polyethylene liners uhmwpe.
The clinical outcome of the multigen plus knee with a biolox delta ceramic femoral component demonstrates results comparable to other studies which reported a mean postoperative 1 year hss score between 85 0 to 93 0 points 20 21.
The ceramic components receive precisely executed final polishing which results in reduced surface roughness around 2 nanometers of all implant materials.
The use of ceramic femoral components in total knee replacement provides tribological and allergological benefits compared to metallic components.
The study was aimed to evaluate the clinical and radiological outcomes of the biolox delta ceramic femoral component and was classified as evidence.
The multigen plus ceramic knee is a non constrained.
Facing the brittleness and the low.
Ceramic multigen plus knee with biolox.
A total of 25 tkas were included in a prospective clinical study in agreement with the inclusion and exclusion criteria after approval of the local ethical committee.
The multigen plus system with a ceramic femoral component lima corporate villanova di san daniele udine italy consists of a biolox delta ceramic femoral component.
All tkrs were implanted in a standardised.
Outcomes and the reliability of the unconstrained multigen plus total knee system with a new biolox delta ceramic femoral component.